6-19
Error Messages
Error Messages in Alphabetical Order
REQUESTED DATA NOT CURRENTLY AVAILABLE
Error Number
30
The analyzer does not currently contain the data you have requested. For
example, this condition occurs when you request error term arrays and no
calibration is active.
SAVE FAILED. INSUFFICIENT MEMORY
Error Number
151
You cannot store an instrument state in an internal register due to insufficient
memory. Increase the available memory by clearing one or more save/recall
registers and pressing
, or by storing files to a disk.
SEGMENT #n POWER OUTSIDE RANGE LIMIT
Information
Message
The selected power range does not support the power level of one or more
segments in the swept list table. This message appears when swept list mode is
not on and reports the first segment that is out of range. Change the segment
power or change the power range.
SEGMENT #n START FREQ OVERLAPS PREVIOUS SEGMENT
Information
Message
A segment entered in the swept list table caused one or more frequency
segments to overlap. This message appears when swept list mode is not on and
reports the first segment that is overlapping another. Change the frequency
ranges of the overlapping segments.
SELECTED SEQUENCE IS EMPTY
Error Number
124
The sequence you attempted to run does not contain instrument commands.
SELF TEST #n FAILED
Service Error
Number 112
Internal test #n has failed. Several internal test routines are executed at
instrument preset. The analyzer reports the first failure detected. Refer to the
service guide for troubleshooting information on internal tests and the
self-diagnose feature.
SEQUENCE ABORTED
Error Number
157
The sequence running was stopped prematurely when you pressed the
key.
SEQUENCE MAY HAVE CHANGED, CAN'T CONTINUE
Error Number
153
When you pause a sequence, you cannot continue it if you have modified it. You
must start the sequence again.
Preset
Local
Summary of Contents for 8719ES
Page 6: ...vi ...
Page 10: ...Contents x Contents ...
Page 11: ...1 1 1 HP 8719 20 22ES Specifications and Characteristics ...
Page 60: ...1 50 HP 8719 20 22ES Specifications and Characteristics Instrument Specifications ...
Page 61: ...2 1 2 HP 8719 20 22ET Specifications and Characteristics ...
Page 98: ...2 38 HP 8719 20 22ET Specifications and Characteristics Instrument Specifications ...
Page 99: ...3 1 3 Front Rear Panel ...
Page 111: ...4 1 4 Menu Maps ...
Page 113: ...4 3 Menu Maps Menu Maps Figure 4 2 Menu Map for Copy ...
Page 114: ...4 4 Menu Maps Menu Maps Figure 4 3 Menu Map for Display ...
Page 115: ...4 5 Menu Maps Menu Maps Figure 4 4 Menu Map for Format Figure 4 5 Menu Map for Local ...
Page 116: ...4 6 Menu Maps Menu Maps Figure 4 6 Menu Map for Marker Marker Fctn and Marker Search ...
Page 118: ...4 8 Menu Maps Menu Maps Figure 4 9 Menu Map for Power and Sweep Setup ET only ...
Page 119: ...4 9 Menu Maps Menu Maps Figure 4 10 Menu Map for Power and Sweep Setup ES only ...
Page 120: ...4 10 Menu Maps Menu Maps Figure 4 11 Menu Map for Preset ...
Page 121: ...4 11 Menu Maps Menu Maps Figure 4 12 Menu Map for Save Recall ...
Page 122: ...4 12 Menu Maps Menu Maps Figure 4 13 Menu Map for Scale Ref ...
Page 128: ...5 1 5 Hardkey Softkey Reference ...
Page 194: ...6 1 6 Error Messages ...
Page 222: ...7 1 7 Options and Accessories ...
Page 234: ...8 1 8 Preset State and Memory Allocation ...
Page 253: ...8 20 Preset State and Memory Allocation Memory Allocation ...
Page 254: ...9 1 9 Understanding the CITIfile Data Format ...
Page 269: ...9 16 Understanding the CITIfile Data Format Useful Calculations ...
Page 270: ...10 1 10 Determining System Measurement Uncertainties ...
Page 281: ...10 12 Determining System Measurement Uncertainties Measurement Uncertainty Equations ...