6-17
Error Messages
Error Messages in Alphabetical Order
POSSIBLE FALSE LOCK
Error Number
6
Phase lock has been achieved, but the source may be phase locked to the wrong
harmonic of the synthesizer. Perform the source pretune correction routine
documented in the "Adjustments and Correction Constants" chapter in the
service guide.
POWER SUPPLY HOT!
Error Number
21
The temperature sensors on the A8 post-regulator assembly have detected an
over-temperature condition. The power supplies regulated on the post-regulator
have been shut down.
POWER SUPPLY SHUT DOWN!
Error Number
22
One or more supplies on the A8 post-regulator assembly have been shut down
due to an over-current, over-voltage, or under-voltage condition.
POWER UNLEVELED
Error Number
179
There is either a hardware failure in the source or you have attempted to set the
power level too high. The analyzer allows the output power to be set higher or
lower than the specified available power range. However, these output powers
may be unleveled or unavailable. Check to see if the power level you set is
within specifications. If it is, refer to the service guide for troubleshooting.
PRINT ABORTED
Error Number
25
When you press the
key, the analyzer aborts output to the printer.
print color not supported with EPSON
Error Number
178
You have defined the printer type as EPSON-P2. Color print is not supported
with this printer. The print will abort.
PRINTER: busy
Error Number
176
The parallel port printer is not accepting data.
PRINTER: error
Error Number
175
The parallel port printer is malfunctioning. The analyzer cannot complete the
copy function.
PRINTER: not connected
Error Number
173
There is no printer connected to the parallel port.
Local
Summary of Contents for 8719ES
Page 6: ...vi ...
Page 10: ...Contents x Contents ...
Page 11: ...1 1 1 HP 8719 20 22ES Specifications and Characteristics ...
Page 60: ...1 50 HP 8719 20 22ES Specifications and Characteristics Instrument Specifications ...
Page 61: ...2 1 2 HP 8719 20 22ET Specifications and Characteristics ...
Page 98: ...2 38 HP 8719 20 22ET Specifications and Characteristics Instrument Specifications ...
Page 99: ...3 1 3 Front Rear Panel ...
Page 111: ...4 1 4 Menu Maps ...
Page 113: ...4 3 Menu Maps Menu Maps Figure 4 2 Menu Map for Copy ...
Page 114: ...4 4 Menu Maps Menu Maps Figure 4 3 Menu Map for Display ...
Page 115: ...4 5 Menu Maps Menu Maps Figure 4 4 Menu Map for Format Figure 4 5 Menu Map for Local ...
Page 116: ...4 6 Menu Maps Menu Maps Figure 4 6 Menu Map for Marker Marker Fctn and Marker Search ...
Page 118: ...4 8 Menu Maps Menu Maps Figure 4 9 Menu Map for Power and Sweep Setup ET only ...
Page 119: ...4 9 Menu Maps Menu Maps Figure 4 10 Menu Map for Power and Sweep Setup ES only ...
Page 120: ...4 10 Menu Maps Menu Maps Figure 4 11 Menu Map for Preset ...
Page 121: ...4 11 Menu Maps Menu Maps Figure 4 12 Menu Map for Save Recall ...
Page 122: ...4 12 Menu Maps Menu Maps Figure 4 13 Menu Map for Scale Ref ...
Page 128: ...5 1 5 Hardkey Softkey Reference ...
Page 194: ...6 1 6 Error Messages ...
Page 222: ...7 1 7 Options and Accessories ...
Page 234: ...8 1 8 Preset State and Memory Allocation ...
Page 253: ...8 20 Preset State and Memory Allocation Memory Allocation ...
Page 254: ...9 1 9 Understanding the CITIfile Data Format ...
Page 269: ...9 16 Understanding the CITIfile Data Format Useful Calculations ...
Page 270: ...10 1 10 Determining System Measurement Uncertainties ...
Page 281: ...10 12 Determining System Measurement Uncertainties Measurement Uncertainty Equations ...