36.
Gain
Compression,
HP
8594EM
6.
On
the
EMC
analyzer
,
press
4
PRESET
5 ,
then
wait
for
the
preset
routine
to
nish.
Press
the
EMC
analyzer
keys
as
follows:
4
FREQUENCY
5
2.0
4
GHz
5
4
SP
AN
5
20
4
MHz
5
4
AMPLITUDE
5
NNNNNNNNNNNNNNNNNNNNNNN
REF
LVL
77
4
dB V
5
N
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
SCALE
LOG
LIN
(LOG)
1
4
dB
5
4
BW
5
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
IF
BW
AUTO
MAN
(MAN)
300
4
kHz
5
7.
On
synthesized
sweeper
#1,
adjust
the
power
level
for
a
0
dBm
reading
on
the
measuring
receiver
.
Set
RF
to
o.
The
power
level
applied
to
the
EMC
analyzer
input
is
10
dB
greater
than
the
specication
to
account
for
the
10
dB
attenuation
setting.
A
power
level
of
0
dBm
at
the
EMC
analyzer
input
yields
010
dBm
at
the
input
mixer
.
8.
Disconnect
the
power
sensor
from
the
directional
coupler
and
connect
the
directional
coupler
to
the
INPUT
50
connector
of
the
EMC
analyzer
using
an
adapter
.
Do
not
use
a
cable
.
9.
On
the
EMC
analyzer
,
press
the
following
keys:
4
MKR
!
5
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
MKR
!
HIGH
4
MKR
5
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
More
1
of
3
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
MK
TRACK
ON
OFF
(ON)
4
SP
AN
5
10
4
MHz
5
W
ait
for
the
AUTO
ZOOM
routine
to
nish.
10.
On
synthesized
sweeper
#2,
adjust
the
power
level
to
place
the
signal
1
dB
below
the
EMC
analyzer
reference
level.
11.
On
the
EMC
analyzer
,
press
4
MKR
!
5
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
MKR
!
HIGH
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
More
1
of
3
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
More
2
of
3
NNNNNNNNNNNNNNNNNNNNNNNNNN
MARKER
1
.
12.
On
synthesized
sweeper
#1,
set
RF
to
ON.
13.
On
the
EMC
analyzer
,
press
4
MKR
!
5
N
NNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNNN
MKR
!
HIGH
,
then
N
NNNNNNNNNNNNNNNNNNNNNNNNNNNN
NEXT
PEAK
.
The
active
marker
should
be
on
the
lower
amplitude
signal
and
not
on
the
signal
that
is
o
the
top
of
the
screen.
If
it
is
not
on
the
lower
amplitude
signal,
reposition
the
marker
to
this
peak
using
the
EMC
analyzer
knob
.
2-250
P
erformance
V
erication
T
ests
Summary of Contents for 8590 E-Series
Page 11: ......
Page 24: ...Calibrating 1 13 ...
Page 29: ......
Page 78: ...10 Residual FM HP 8591EM Slope Hz dB Performance Veri cation Tests 2 49 ...
Page 431: ......
Page 432: ...3 Performance Test Records Performance Test Records 3 1 ...
Page 433: ...3 2 Performance Test Records ...
Page 446: ...HP 8591EM Performance Test Record Performance Test Records 3 15 ...
Page 461: ...HP 8593EM Performance Test Record 3 30 Performance Test Records ...
Page 474: ...HP 8594EM Performance Test Record Performance Test Records 3 43 ...
Page 487: ...HP 8595EM Performance Test Record 3 56 Performance Test Records ...
Page 501: ......
Page 528: ...Regulatory Information HP 8591EM Speci cations and Characteristics 4 27 ...
Page 547: ...Amplitude Characteristics DYNAMIC RANGE 5 18 HP 8593EM Speci cations and Characteristics ...
Page 560: ...Regulatory Information HP 8593EM Speci cations and Characteristics 5 31 ...
Page 589: ...Regulatory Information 6 28 HP 8594EM Speci cations and Characteristics ...
Page 591: ......
Page 619: ...Regulatory Information 7 28 HP 8595EM Speci cations and Characteristics ...
Page 621: ......
Page 650: ...Regulatory Information HP 8596EM Speci cations and Characteristics 8 29 ...
Page 675: ......
Page 684: ...Customer Support 10 9 ...
Page 685: ...10 10 Customer Support ...
Page 693: ......