51.
P
ower
Sweep
Range
,
HP
8593EM
,
HP
8594EM
,
HP
8595EM
,
and
HP
8596EM
This
performance
verication
test
must
be
performed
with
the
EMC
analyzer
set
in
the
spectrum
analyzer
mode.
The
tracking
generator
output
is
connected
to
the
EMC
analyzer
INPUT
50
through
a
power
splitter
and
the
tracking
is
adjusted
at
300
MHz
for
a
maximum
signal
level.
The
other
output
of
the
power
splitter
is
connected
to
a
measuring
receiver
.
The
tracking
generator
is
set
to
do
a
power
sweep
from
97
dBV
to
106
dBV
(010
dBm
to
01
dBm).
The
markers
are
used
to
measure
the
displayed
amplitude
at
the
beginning
and
end
of
the
sweep
.
The
power
sweep
is
then
turned
o
and
the
power
level
of
the
tracking
generator
is
adjusted
until
the
displayed
amplitude
is
the
same
as
at
the
start
of
the
sweep
.
This
power
level
is
measured
on
the
measuring
receiver
and
recorded.
The
tracking
generator
is
then
adjusted
until
the
displayed
amplitude
is
the
same
as
at
the
end
of
the
sweep
.
This
power
level
is
measured
and
recorded.
The
dierence
between
the
two
measured
power
levels
is
calculated
and
recorded.
The
related
adjustment
for
this
performance
verication
test
is
the
\Tracking
Generator
P
ower
Level
for
Option
010,
HP
8593EM,
HP
8594EM ,
HP
8595EM,
and
HP
8596EM."
Equipment
Required
Measuring
receiver
P
ower
sensor
,
100
kHz
to
2.9
GHz
P
ower
splitter
Cable
,
Type
N,
62
cm
(24
in)
A
dapter
,
Type
N
(m)
to
Type
N
(m)
2-330
P
erformance
V
erication
T
ests
Summary of Contents for 8590 E-Series
Page 11: ......
Page 24: ...Calibrating 1 13 ...
Page 29: ......
Page 78: ...10 Residual FM HP 8591EM Slope Hz dB Performance Veri cation Tests 2 49 ...
Page 431: ......
Page 432: ...3 Performance Test Records Performance Test Records 3 1 ...
Page 433: ...3 2 Performance Test Records ...
Page 446: ...HP 8591EM Performance Test Record Performance Test Records 3 15 ...
Page 461: ...HP 8593EM Performance Test Record 3 30 Performance Test Records ...
Page 474: ...HP 8594EM Performance Test Record Performance Test Records 3 43 ...
Page 487: ...HP 8595EM Performance Test Record 3 56 Performance Test Records ...
Page 501: ......
Page 528: ...Regulatory Information HP 8591EM Speci cations and Characteristics 4 27 ...
Page 547: ...Amplitude Characteristics DYNAMIC RANGE 5 18 HP 8593EM Speci cations and Characteristics ...
Page 560: ...Regulatory Information HP 8593EM Speci cations and Characteristics 5 31 ...
Page 589: ...Regulatory Information 6 28 HP 8594EM Speci cations and Characteristics ...
Page 591: ......
Page 619: ...Regulatory Information 7 28 HP 8595EM Speci cations and Characteristics ...
Page 621: ......
Page 650: ...Regulatory Information HP 8596EM Speci cations and Characteristics 8 29 ...
Page 675: ......
Page 684: ...Customer Support 10 9 ...
Page 685: ...10 10 Customer Support ...
Page 693: ......