11.
Residual
FM,
HP
8593EM
,
HP
8594EM
,
HP
8595EM
,
and
HP
8596EM
11.
Residual
FM,
HP
8593EM
,
HP
8594EM
,
HP
8595EM
,
and
HP
8596EM
This
performance
verication
test
must
be
performed
with
the
EMC
analyzer
set
in
the
spectrum
analyzer
mode.
This
test
measures
the
inherent
short-term
instability
of
the
EMC
analyzer
LO
system.
With
the
analyzer
in
zero
span,
a
stable
signal
is
applied
to
the
input
and
slope-detected
on
the
linear
portion
of
the
IF
bandwidth
lter
skirt.
Any
instability
in
the
LO
transfers
to
the
IF
signal
in
the
mixing
process
.
The
test
determines
the
slope
of
the
IF
lter
in
hertz
per
decibel
(Hz/dB)
and
then
measures
the
signal
amplitude
variation
caused
by
the
residual
FM.
Multiplying
these
two
values
yields
the
residual
FM
in
hertz.
The
narrow
bandwidth
test
uses
a
300
Hz
span.
This
span
is
not
specied,
however
,
it
is
tested
in
\Frequency
Span
A
ccuracy
."
There
are
no
related
adjustment
procedures
for
this
performance
test.
Equipment
Required
Signal
generator
Cable
,
Type
N,
183
cm
(72
in)
A
dditional
Equipment
for
Option
026
A
dapter
,
APC
3.5
(f)
to
Type
N
(f)
Procedure
This
performance
verication
test
consists
of
two
parts:
P
art
1:
Residual
FM
P
art
2:
Narrow
Bandwidth
Residual
FM
P
erformance
V
erication
T
ests
2-51
Summary of Contents for 8590 E-Series
Page 11: ......
Page 24: ...Calibrating 1 13 ...
Page 29: ......
Page 78: ...10 Residual FM HP 8591EM Slope Hz dB Performance Veri cation Tests 2 49 ...
Page 431: ......
Page 432: ...3 Performance Test Records Performance Test Records 3 1 ...
Page 433: ...3 2 Performance Test Records ...
Page 446: ...HP 8591EM Performance Test Record Performance Test Records 3 15 ...
Page 461: ...HP 8593EM Performance Test Record 3 30 Performance Test Records ...
Page 474: ...HP 8594EM Performance Test Record Performance Test Records 3 43 ...
Page 487: ...HP 8595EM Performance Test Record 3 56 Performance Test Records ...
Page 501: ......
Page 528: ...Regulatory Information HP 8591EM Speci cations and Characteristics 4 27 ...
Page 547: ...Amplitude Characteristics DYNAMIC RANGE 5 18 HP 8593EM Speci cations and Characteristics ...
Page 560: ...Regulatory Information HP 8593EM Speci cations and Characteristics 5 31 ...
Page 589: ...Regulatory Information 6 28 HP 8594EM Speci cations and Characteristics ...
Page 591: ......
Page 619: ...Regulatory Information 7 28 HP 8595EM Speci cations and Characteristics ...
Page 621: ......
Page 650: ...Regulatory Information HP 8596EM Speci cations and Characteristics 8 29 ...
Page 675: ......
Page 684: ...Customer Support 10 9 ...
Page 685: ...10 10 Customer Support ...
Page 693: ......