Setting Global System Parameters
How to Set the Characteristics of the Trigger Output
92
HP 81200 Data Generator/Analyzer Platform User Guide, Revision 2.1
How to Set the Characteristics of
the Trigger Output
The TRIGGER OUTPUT connector of the central clock module can be
used to inform other instruments upon an event. This event can be:
•
Start of a data block within the overall test sequence
•
An event detected by the system (see
“Event Handling Principles” on
).
To set the trigger characteristics:
1
Open the Parameter Editor for the master clock module (see
Start the Parameter Editor for Global Parameters” on page 85
The Clock Reference window appears.
2
Click the
Trigger Output
tab.
Figure 30
Trigger Output Configuration Window
3
Set the voltages and output impedance.
4
Select the Mode.
–
Clock Generator
: generates a clock signal derived from the system
clock.
–
Sequencer
: Generates a transition from high to low or vice versa.
5
Set the delay, if desired. The delay refers to the system clock.
6
Frequency Multiplier
is only available in Clock Generator mode. You
can choose one of the factors from the list. The list shows the FMR (see
Summary of Contents for 81200
Page 1: ...Revision 2 1 HP 8 1 200 D a t a G e n e r a t o r A n a l y z e r P l a t f o r m User Guide ...
Page 93: ...HP 81200 Data Generator Analyzer Platform User Guide Revision 2 1 93 Connecting the DUT ...
Page 181: ...HP 81200 Data Generator Analyzer Platform User Guide Revision 2 1 181 Running the Test ...
Page 215: ...HP 81200 Data Generator Analyzer Platform User Guide Revision 2 1 215 How Do I ...
Page 240: ...Index 240 HP 81200 Data Generator Analyzer Platform User Guide Revision 2 1 ...