How to Set Up a Port Or Channel for Signals To Be Analyzed
Setting Up Ports and Channels
HP 81200 Data Generator/Analyzer Platform User Guide, Revision 2.1
123
The settings of
Input Type
and
Polarity
have the following effects on
data recognition:
4
Activate the connection. This is probably the most important step of all!
T I P
When running a test, check the green LEDs of the frontends. They indicate
whether the channel is enabled or disabled.
5
Close the window.
Table 5
Data Acquisition With HP E4837A Differential Analyzer
Input Type
Input Signal Condition
Acquired Data
Normal Polarity
Inverted Polarity
Differential
Signal at IN > IN
1
0
Signal at IN < IN
0
1
Normal Input
Signal at IN > Threshold
1
0
Signal at IN < Threshold
0
1
Complementary Input
Signal at IN > Threshold
0
1
Signal at IN < Threshold
1
0