How to Set Up a Port or Channel for Generated Signals
Setting Up Ports and Channels
HP 81200 Data Generator/Analyzer Platform User Guide, Revision 2.1
111
–
Hold Width
or
Hold Duty Cycle
: Set a preference if the system clock
frequency is changed.
–
Transition times
(with HP E4842A or E4838A frontends only):
The Rise/Fall Time can be set in the range of 0.5 to 10.0 ns to achieve
the specified 0.7 to 6.0 ns. The rise and fall times are coupled and
yield a signal as shown below:
Figure 38
Variable Slopes Setting
–
Format
: Choose from the list.
Figure 39
Signal Formats
The default format for a data port and the channels connected to a data
port is NRZ. The default format for a pulse port or an unconnected
generator frontend is RZ.
NRZ
RZ
R1
1
0
1
1
Non-return to zero
Return to zero
Return to one