How Do I ... ?
How Can I Generate a Clock Signal of up to 1.3 GHz?
220
HP 81200 Data Generator/Analyzer Platform User Guide, Revision 2.1
4
Use the Parameter Editor to adjust the characteristics of the upper
channel. Keep the RZ format. From the
Frequency Multiplier
field,
choose the desired frequency (a 2
n
multiple or fraction of the system
clock frequency).
Figure 121
Parameter Editor Window
In this example we multiply the system clock frequency of 100 MHz by a
factor of two. This yields a clock period of 5 ns. Do not forget to switch
the generator output on.
5
Use the Parameter Editor to adjust the characteristics of the lower
channel. Keep the RZ format. Set a half-period delay.
Figure 122
Parameter Editor Window