CONFIGURATION AND ANALYSIS SOFTWARE ENSUITE
Information for general use
Rev. M / 73023639
166
GasLab Q2
8.3.4.
Viewing position of the security switch
Some settings must be protected as mentioned above for fiscal measurements.
This is done using a physical connection (jumper) inside the device, known as the security switch
(SSW), compare section
6.1.3 Opening and closing the hardware parameter guard (SSW)
Independent of the use in fiscal measurements, the administrator has the possibility to place further
device settings under this protection.
You can view the setting using
online parameterization
and the online help of
enSuite).
After opening the online parameterization, select the elements as shown in the figure.
Figure 8.20: View the SSW position (open and closed)
On delivery and in “non
-
fiscal operation”
,
this connection is open.
When the switch is closed, the text
"
The security switch is currently closed
" is displayed.