GasAlertMicro 5/PID/IR
User Manual
18
Bump Daily Enabled
a
Caution
BW recommends that a bump test to all sensors be
performed every 24 hours prior to the beginning of the
work shift.
13. If
Bmp Daily
(bump daily) is enabled in tech mode, the follow-
ing screens display.
If a bump test of the LEL and O
2
sensor is not performed, the
detector will deactivate.
Apply the LEL gas and then apply the O
2
(a higher or lower
percentage than the default 20.9%, such as 18% O
2
).
Successful Bump Test:
If the bump test passes, the following screens
display.
The detector waits for the sensor(s) to clear (30 seconds) and then
enters normal operation.
Unsuccessful Bump Test:
If the bump test is unsuccessful or the bump
test is not performed, the following screen displays and the detector
deactivates.