GasAlertMicro 5/PID/IR
User Options Menu
LEL Sensor
The Custom LEL correction screen displays. Press
G
or
H
to select the
required value, and press
C
within 5 seconds to confirm.
PID Sensor
This option is used to enter compensation
factors for hydrocarbons other than
methane. The factor can only be applied if
the LEL sensor has been calibrated with
methane.
After selecting the
LEL
sensor, press
H
to scroll to
Correction
. Press
C
within
20 seconds to confirm and access the LEL
correction library.
Scroll to the required gas type and
press
C
. A check displays in the
corresponding checkbox. The detector
automatically applies the correction factor.
To disable the
Correction
option, press
H
to scroll to
None
or to
Methane
. A check
displays in the corresponding checkbox. If
required, select a different gas type
correction factor.
Custom:
To enter a correction factor that
is not listed in the library, press
H
to scroll
to
Custom
and press
C
within 5 seconds
to confirm.
This option is used to enter compensation
factors for selected gas types. The factor
can only be applied if the PID sensor has
been calibrated with isobutylene.
After selecting the
PID
sensor, press
H
to scroll to
Correction
. Press
C
within
20 seconds to confirm and access the
PID correction library.
Scroll to the required gas type and
press
C
. A check displays in the
corresponding checkbox. The detector
automatically applies the correction factor.
To disable the
Correction
option,
press
H
to scroll to
None
or to
Isobutyl
.
A check displays. If required, select a
different gas type correction factor.
Custom:
To enter a correction factor for
a custom PID sensor, scroll to
Custom
and press
C
. Press
G
or
H
to scroll to
the required value, and press
C
within
5 seconds to confirm. Refer to the PID
Correction Factor
Library for gas types and
corresponding correction factor values.