17
GasAlertMicro 5/PID/IR
Activating the Detector
If any sensor is over due for calibration, the LCD displays the
name of the sensor and the number of days past due.
Due-Lock Enabled
The
Due-lock
option ensures that a passcode must be entered when
calibration is past due, otherwise the detector automatically deactivates.
11. If
Due-Lock
is enabled in the user options, the following
screen displays.
Enter the correct passcode and press
C
to confirm.
Note
If any sensor is overdue,
Y
displays continually until calibration
is performed.
If no passcode is entered, or it is entered incorrectly, the follow-
ing screen displays.
To enable/disable this option, refer to
in the user
options menu. Also refer to
.
Force Calibration Enabled
12. If
Force cal
(force calibration) is enabled in tech mode, calibra-
tion is mandatory before the detector enters normal operation.
in
to enable/disable, and
refer to
to calibrate.
If
C
is not pressed to start calibration, the following screen dis-
plays and the detector deactivates.