GasAlertMicro 5/PID/IR
Specifications
Specifications
Instrument dimensions:
14.5 x 7.4 x 3.8 cm (5.7 x 2.9 x 1.5 in.)
Weight:
370 g (13.1 oz.)
Operating and storage conditions
Operating Temperature:
VOC: -10°C to +40°C (-14°F to +104°F)
Other gases: -20°C to +50°C (-4°F to +122°F)
Combustible gas sensor: Certified by CSA International to ±3% LEL
accuracy from -10°C to +40°C (4°F to 104°F)
Storage temperature:
-25°C to +60°C
Humidity:
O
2
: 0% to 99% relative humidity (non-condensing)
VOC and CO
2
IR: 0% to 95% relative humidity (non-condensing)
Combustibles: 5% to 95% relative humidity (non-condensing)
Cl
2
: 10% to 95% relative humidity (non-condensing)
HCN, ClO
2
: 15% to 95% relative humidity (non-condensing)
Other gases: 15% to 90% relative humidity (non-condensing)
Pressure:
95 to 110 kPa
Alarm setpoints:
May vary by region and are user-defined
Detection range:
O
2
: 0 – 30.0% vol. (0.1% vol. increments)
CO: 0 – 999 ppm (1 ppm increments)
CO (TwinTox sensor): 0 – 500 ppm (1 ppm increments)
H
2
S: 0 – 500 ppm (1 ppm increments)
H
2
S (TwinTox sensor): 0 – 500 ppm (1 ppm increments)
Combustible (LEL): 0 - 100% LEL (1% LEL increments) or 0 - 5.0% v/v
methane; certified by CSA International to C22.2 No. 152 and ISA
12.13.01 within 0 - 60% or 3.0% v/v methane
PH
3
: 0 – 5.0 ppm (0.1 ppm increments)
SO
2
: 0 – 150 ppm (1 ppm increments)
Cl
2
: 0 – 50.0 ppm (0.1 ppm increments)
NH
3
: 0 – 100 ppm (1 ppm increments)
NO
2
: 0 – 99.9 ppm (0.1 ppm increments)
HCN: 0 – 30.0 ppm (0.1 ppm increments)
ClO
2
: 0 – 1.00 ppm (0.01 ppm increments)
O
3
: 0 – 1.00 ppm (0.01 ppm increments)
VOC: 0 – 1000 ppm (1.0 ppm increments)
CO
2
IR: 0-50,000 ppm (50 ppm increments) or 0-5.0% v/v CO
2
(Sensors not available for use with the GasAlertMicro 5 IR: ClO
2
, HCN,
NO
2
, PH
3
, and Cl
2
)
Sensor type:
H
2
S/CO: Twin plug-in electrochemical cell
Combustibles: Plug-in catalytic bead
VOC: Photoionization detector (PID)
CO
2
: IR detector
Other gases: Single plug-in electrochemical cell
O
2
measuring principle:
Capillary controlled concentration sensor
Alarm conditions:
TWA alarm, STEL alarm, low alarm, high alarm,
multi alarm, over limit alarm, sensor alarm, pump alarm,
MMC/SD card fail alarm, low battery alarm, confidence beep,
automatic deactivation alarm
Audible alarm:
95 dB at 0.3 m (1 ft.) variable pulsed dual beepers