29
GasAlertMicro 5/PID/IR
User Options Menu
Resolution
Note
Regular
resolution for O
3
and ClO
2
sensors is 0.1 ppm, while
extra resolution is 0.01 ppm.
The
Resolution
option is not available for CO, O
2
, LEL, PID,
and CO
2
sensors.
After selecting the desired sensor, press
H
to scroll to
Resolution
.
Press
C
within 20 seconds to confirm.
%Vol CO
2
(CO
2
Sensors Only)
Note
Correction factors are not applicable to the CO
2
sensor.
The
Resolution
option displays the gas
measurement using
Regular
or
Extra
resolution.
Regular:
Displays gas measurement in
1 ppm increments.
Extra:
Displays the gas measurement in
0.1 ppm increments.
A check displays in the checkbox of the
currently selected resolution. To select the
other resolution, press
H
to move the
check to other resolution. Press
C
to
confirm the selection.
The detector is shipped with
Regular
resolution enabled for applicable sensors.
If the
%vol
CO
2
is enabled, the detector
displays the carbon dioxide (CO
2
)
readings as %vol (
0.0
).
From the
Sensors
option menu, select
CO
2
.
Press
H
to scroll to
%vol CO
2
. Press
C
to toggle between enable and disable.
Confirmation is not required. If no buttons
are pushed, after 20 seconds the detector
returns to the sensor selection screen.
The change is saved automatically.
The detector is shipped with
%vol CO
2
disabled.