GasAlertMicro 5/PID/IR
User Manual
58
Note
If a value is changed but
C
is not pressed within 5 seconds to
confirm, the following screen displays.
The previous value is automatically retained. The detector
proceeds to the next sensor due date.
Repeat step #7 to set the calibration due dates for the remain-
ing sensors.
When all of the due dates are entered, press
C
to proceed to
Or
Press
A
to bypass setting the alarm setpoints and proceed to
step #10.
Alarm Setpoints
8. When all of the sensor due dates have been set or bypassed,
the alarm setpoints need to be set or bypassed. The following
screen displays.
Factory alarm setpoints may vary by region. Refer to
for an example.
Alarms can be set to any value within the detection range of
the selected sensor. Refer to Detection Range in
.
Note
To disable an alarm setpoint, set it to
0
(zero).