GasAlertMicro 5/PID/IR
User Manual
36
Sensors
a
Caution
To reconfigure the sensor type, physically change the
sensor prior to entering Tech mode.
When a toxic sensor is physically removed and replaced by another toxic
sensor, the detector must be reconfigured to recognize the change.
Note
If a sensor is replaced, the detector will classify the sensor as
overdue for calibration. Calibrate the new sensor immediately.
1.
Press
H
to scroll to
Sensors
. Press
C
within 20 seconds to confirm and
access the toxic sensor menu.
2.
Press
G
or
H
to scroll to
Toxic 1
or
Toxic
2
and press
C
within
20 seconds to confirm.
A corresponding list of toxic sensors
displays. A checkbox displays beside
the current toxic sensor.
Note
Toxic 1:
List includes the PID and CO
2
sensors.
Toxic 2:
List includes the H
2
S/CO COSH
sensor.
3.
Press
G
or
H
to scroll to the new
sensor and press
C
to confirm. A
checkbox displays beside the new
sensor. To reconfigure, exit the user
options menu.
The following screen displays. The
detector deactivates and immediately
reactivates. It performs the
reconfiguration during the startup.
The new sensor must be calibrated.
Until the new sensor is calibrated, the
due date displays as
OL
(over limit)
when in normal operation.
4.
Calibrate the new sensor
immediately.