GasAlertMicro 5/PID/IR
User Manual
38
Bump Daily
Note
If
Bmp daily
is enabled, the O
2
and LEL sensors require daily
bump tests whenever the detector is activated (or reactivated)
following 00:00 hours (midnight).
Stealth Mode
The detector is shipped with the
Stealth
option disabled.
If enabled, the
Bmp
daily
option forces the
detector to perform a daily bump test to
ensure that it is responding to the test gas.
Press
H
to scroll to
Bmp daily
. Press
C
to toggle between enable/disable.
When enabled, the following screen
displays during the startup self-test.
The bump test must be performed for the
LEL and O
2
sensors, otherwise the
detector will deactivate. Refer to
for procedures.
The detector is shipped with the
Bmp
daily
option disabled.
The
Stealth
option disables the backlight,
visual alarms, and audible alarms when
concealment is required.
Only the vibrator and the LCD activate
during an alarm condition.
Press
H
to scroll to
Stealth
. Press
C
to
toggle between enable and disable.