GasAlertMicro 5/PID/IR
User Manual
60
Setting the Low Alarm Setpoint
The current low alarm setpoint displays for the selected sensor.
Press
H
or
G
to change the value for the low alarm setpoint. When the
required value displays, press
C
to confirm.
Setting the High Alarm Setpoint
The current high alarm setpoint displays for the selected sensor.
Press
H
or
G
to change the value for the high alarm setpoint. When the
required value displays, press
C
to confirm.
Proceed to step #9
Setting the Remaining Alarm Setpoints
.
Setting the Remaining Alarm Setpoints
9. Repeat step #8 to set the alarm setpoints for the remaining
sensors. When complete, the detector emits two quick beeps
and proceeds to
Finish Calibration
10. The detector displays the following screen to indicate that
calibration is complete.