36
Annual Inspection
en
1. Switch system on. (See Chapter 5.4, Turning On the Aquilex System)
2. Insert tube set into pump and close clamps on the bags.
3. Hang the fluid bags onto the fluid bag hooks.
4. Insert tap spikes into the fluid bags and open clamps on the fluid bags.
5. Insert hysteroscope tube into the measuring cup.
6. Set intrauterine nominal pressure to 150 mmHg.
7. Press the
Prime
button
8. The roller wheel starts to rotate in order to remove the air from the tubes and
to carry out the automatic lumen calibration.
9. When the automatic lumen calibration is completed, press
Pause/Resume
button (about 20 seconds).
10. Empty the measuring cup.
11. Place hysteroscope tube back into the measuring cup.
12. Press the
Pause/Resume
button.
13. Press
Pause/Resume
button after one minute. The measuring cup should
contain approx. 800 ml +/- 60 ml of fluid.
Enter the results into the test log in section 16.1. The test is successfully complet-
ed when the results are within the permissible tolerance limit.
9.2.3
Pressure Measuring Test
The test setup is depicted in Fig. 9-2.
The pressure test checks the pressure chamber, the pressure sensor, and the ac-
curacy of the pressure measurement, to ensure that all elements function cor-
rectly. For this test, an inflow tube set and a
water-filled
container is required.
The height of the water column (hydrostatic pressure) is used to test the pressure
transducer.
Fig. 9-2
Set-up of pressure measuring
test
h
Height of the water line
Summary of Contents for Aquilex
Page 4: ......
Page 8: ......
Page 10: ......
Page 61: ...Appendix 53 en 16 Appendix 16 1 Test Log Date Tests Performed Results Comment Signature ...
Page 64: ......
Page 120: ......
Page 122: ......
Page 178: ......
Page 180: ......
Page 236: ......
Page 238: ......
Page 294: ......
Page 296: ......
Page 353: ......
Page 354: ......