System Operation
23
en
5.8
Connecting the Outflow Tube Set
CAUTION!
When using the Aquilex System with tissue removal systems, e.g. MyoSure®, the
combination of low set pressures and excessive vacuum pressures may result in
a significant loss of intrauterine distension pressure which has the potential to
affect the visibility of the surgical field. Conversely, when employing a high dis-
tension pressure, the deactivation of the tissue removal system can lead to pres-
sure spikes that can exceed 150 mmHg.
Using low vacuum configuration of Fig. 5-8, connect outflow tube set (Y-tube) to
patient port
(4)
of second container. Yellow flexible connector attaches to drape
(6)
. Yellow luer fitting connects to stopcock
(7)
of removable outflow channel
or hysteroscope outflow valve.
Fig. 5-8
Outflow tube set
(1)
To low vacuum port (white)
(2)
Container
(3)
Connecting tube
(4)
Patient port
(5)
Outflow tube set
(6)
Drape
(7)
Removable outflow channel or hys-
teroscope outflow sheath stopcock
(2)
(6)
(3)
(4)
(5)
(1)
(7)
(5)
(3)
(1)
Summary of Contents for Aquilex
Page 4: ......
Page 8: ......
Page 10: ......
Page 61: ...Appendix 53 en 16 Appendix 16 1 Test Log Date Tests Performed Results Comment Signature ...
Page 64: ......
Page 120: ......
Page 122: ......
Page 178: ......
Page 180: ......
Page 236: ......
Page 238: ......
Page 294: ......
Page 296: ......
Page 353: ......
Page 354: ......