32
Care and Maintenance
en
8
Care and Maintenance
Care and maintenance
The service and maintenance of the system and its accessories has to be carried
out as per instructions to ensure the safe operation of the system. For the protec-
tion of the patient and the operating team, check that the system is complete
and functional before each use.
Special care is necessary when servicing, maintaining, and storing the system
and its accessories to maintain the functionality of the equipment and any at-
tached devices.
8.1
Cleaning the System
1. Use the
ON/OFF
switch to turn off the system.
2. Remove the power cord.
3. Wipe the surface of the system with a soft cloth moistened with a disinfec-
tant (for example, Meliseptol® rapid). The concentration of the used disinfec-
tant depends on the information provided by the manufacturer of the
disinfectant. Make sure moisture does not enter the system.
WARNING!
Cleaning the system / Sterilization not allowed.
The pump and the cart/scale can be disinfected by wiping off the outer surfaces.
Do not sterilize the pump and the cart/scale.
8.2
Maintenance Carried out by Authorized Service Technician
Two-year maintenance interval
It is recommended that an authorized service technician inspects and services
the system at appropriate intervals to ensure safety and functionality. The
mini-
mum
service interval is two years, depending on frequency and duration of use.
If this interval is not maintained, the manufacturer does not assume any liability
for the functional safety of the system. A sticker located on the rear panel of the
system contains the latest date for the next service or maintenance check.
NOTE!
Service and maintenance work may not be carried out during surgery.
Certification
Ask the service technician for a certificate after he or she has inspected the sys-
tem or performed any service tasks. This certificate must list:
• type and scope of service,
• date of service,
• name of company performing service
• as well as signature.
Summary of Contents for Aquilex
Page 4: ......
Page 8: ......
Page 10: ......
Page 61: ...Appendix 53 en 16 Appendix 16 1 Test Log Date Tests Performed Results Comment Signature ...
Page 64: ......
Page 120: ......
Page 122: ......
Page 178: ......
Page 180: ......
Page 236: ......
Page 238: ......
Page 294: ......
Page 296: ......
Page 353: ......
Page 354: ......