Error and Warning Messages
41
en
Scale Overload/Check Scale
3 warning beeps
Weight on the
container scale
exceeds
25 kg (55 lbs) or weight on the
bag scale
exceeds 12 kg (26.5 lbs). The weight
must be reduced. System function will
resume when the excess weight has
been removed.
Communication Error
5 warning beeps
Contact Hologic Technical Support.
Calibration Error
5 warning beeps
Contact Hologic Technical Support.
Sensor Error
5 warning beeps
Contact Hologic Technical Support.
Motor Error
5 warning beeps
Contact Hologic Technical Support.
Low Vac Failed Use Alternative
3 warning beeps
An alternative low vacuum source must
be used in order to continue the proce-
dure. Contact Hologic Technical Support.
High Vac Failed Use Alternative
3 warning beeps
An alternative high vacuum source must
be used in order to continue the proce-
dure. Contact Hologic Technical Support.
Vac Systems Out Use Alternative
3 warning beeps
An alternative vacuum source must be
used in order to continue the procedure.
Contact Hologic Technical Support.
Message in the pump display
Warning beeps
Method