![Hologic Aquilex Instructions For Use Manual Download Page 51](http://html.mh-extra.com/html/hologic/aquilex/aquilex_instructions-for-use-manual_2146510051.webp)
Technical Data
43
en
1
A tube set is not an application part in terms of the standard. However, it meets
all the technical requirements for an application part.
Measurement range
Pressure [mmHg]
0-500 mmHg
-995/+9995 ml
Deficit [ml]
Accuracy
Pressure [mmHg]
± 10 mmHg
Deficit [ml]
< 1 l: ± 60 ml
> 1 l: ± 6 %
Total inflow volume [ml]
± 10 %
Dimensions
Width x Height x Depth [mm], [in] 300 mm x 140 mm x 300 mm/11.8 in x 5.5 in x 11.8
in (AQL-100P),
670 mm x 1390 mm x 670 mm/26.4 in x 54.8 in x 26.4 in
(AQL-100CBS)
Weight [kg], [lbs]
5.8 kg, 12.8 lbs (AQL-100P)
14.2 kg, 31.3 lbs (AQL-100CBS)
Interfaces:
ON/OFF signal for components
Irrigation Pump Unit:
1 x scale connection (flanged socket/round connection
socket with 5 pins)
1 x service port (RS232 socket DSUB9/D-SUB9)
Fluid Monitoring Unit/Container Scale:
1 x data connection (flanged female / round socket with 5
pins) for connection to bag scale.
Fluid Monitoring Unit/Bag Scale:
2 x data ports (flanged socket/round connection socket
with 5 pins): Connector to connect to pump, socket to con-
nect with container scale
Mains power socket
IEC 60320-1 C14
RFID Transponder Technology
Working frequency: 13.5609 MHz,
Transmitting power: -7.51 dB
μ
A/m at 10 m
Essential Performance
Pressure build-up in the body cavity, control and measure-
ment, limit: max. 150 mmHg (Normal Condition), 200
mmHg for max. 5 seconds (Single Fault Condition).
Suction pressure: Limit value: min. - 84 kPa
Summary of Contents for Aquilex
Page 4: ......
Page 8: ......
Page 10: ......
Page 61: ...Appendix 53 en 16 Appendix 16 1 Test Log Date Tests Performed Results Comment Signature ...
Page 64: ......
Page 120: ......
Page 122: ......
Page 178: ......
Page 180: ......
Page 236: ......
Page 238: ......
Page 294: ......
Page 296: ......
Page 353: ......
Page 354: ......