20
System Operation
en
5.4
Turning On the Aquilex System
1. Press the ON/OFF switch. The displays and indicators light up and system
turns on.
2. The system now performs a device self-test.
The device self-test is used by the system to check whether the container and
the bag scale are connected, among others. If the bag scale is not connected,
the message "System OK. Bag Scale Not Connected” is displayed. In this case,
please check the appropriate data connection of the scales with the pump ac-
cording to chapter 5.3.
3. If a tube set is in the inflow tube holder when the pump is switched on, the
pump display
(Fig. 5-1, Front of Irrigation Pump Unit
(1)
) shows the message
Remove Tube Set
. The device self-test resumes once the tube set is removed
from the roller wheel.
If the device self-test is unsuccessful, the corresponding error messages are
displayed (see Chapter 10, Error and Warning Messages).
The system has successfully completed the device self-test when a single audible
beep is heard. The message
System OK
is displayed for 5 secs followed by the
message
Insert Tube Set
.
Fig. 5-6
High vacuum tube
(6)
Hygiene filter
(7)
Vacuum tube (green connectors)
(8)
Connection for high vacuum
(green)
(9)
Container
(10)
Tissue trap
(11)
Adjustment dial
(12)
Connecting tube
MyoS
ure
®
only f
or
(7)
(8)
(9)
(6)
(10)
(11)
(12)
Summary of Contents for Aquilex
Page 4: ......
Page 8: ......
Page 10: ......
Page 61: ...Appendix 53 en 16 Appendix 16 1 Test Log Date Tests Performed Results Comment Signature ...
Page 64: ......
Page 120: ......
Page 122: ......
Page 178: ......
Page 180: ......
Page 236: ......
Page 238: ......
Page 294: ......
Page 296: ......
Page 353: ......
Page 354: ......