69
USER MANUAL DO-LBL-06035(L)
| MyoVista® Wavelet ECG (
wav
ECG™) 12-Lead Cardiac Testing Device
Device Settings
4 . Select the desired system update package from the pop-up menu and press OK . File name will
either have the .uph file extension or select by system update text .
The MyoVista Device will validate the update package before the following message appears:
WARNING: Do not turn off the device while the system update is in progress.
Restoring Device to Factory Default Settings
(This is accessible only by Administrator roles .)
The factory reset function resets device and user settings to their factory default settings and
removes
all patient data and all instructional/reference documents
. It does not change the MyoVista Device
software version . During this procedure, you have the option to back up the database .
Warning: Do not remove power from the device until the update/factory reset process
completes.
To restore the MyoVista Device to factory default settings:
1 . Under System Settings, select the Update/Factory Reset icon . The Update/Factory Reset screen
appears (see Figure 5-24) .
2 . Select the Factory Reset icon . The following message appears:
A factory reset will erase the database. Do you want to back it up?