Service – Maintenance & Cleaning
Warnings
Warning: DANGER OF ELECTRIC SHOCK. THERE ARE NO USER-SERVICEABLE PARTS IN THIS DEVICE.
The battery is not user-serviceable. Do not open the device case. Only personnel authorized by the
manufacturer may service the device, open the enclosure, or open the battery compartment.
Note: MyoVista Device is a self calibrating ECG device and requires no user/technician
calibration after manufacture.
Service & Maintenance
Regular maintenance, regardless of how often the MyoVista Device is used, is essential to ensure that
the equipment will be functional when used . Regular maintenance includes cleaning and inspection .
Maintenance Items To Remember
Before performing maintenance on the device, be sure that an authorized representative has
provided proper training of device functions and safety precautions .
If there is any uncertainty about maintenance or cleaning instructions or procedures, refer to the
supplied maintenance and cleaning instructions .
The following maintenance instructions do not override any statutory or local regulations or
procedures for the prevention of accidents and environmental protection .
When inspecting and cleaning the device:
• Turn off the system and unplug the device .
• Do not immerse any part of the equipment in liquid .
• Do not use organic solvents, ammonia-based solutions, or abrasive cleaning agents . These may
damage the equipment surfaces .
Warning: Before cleaning, ensure power is turned off and that the power adapter is unplugged.
Caution: DO NOT use corrosive cleaning agents. DO NOT use cleaning agents that contain phenols.
Caution: DO NOT use detergent solutions for cleaning equipment or cable accessories. DO NOT
immerse equipment and cables in liquid.