Troubleshooting
Consistent Environmental Noise in
Signal
AC line power EMI
Environmental EMI (OTA) – Seen as
elevated when patient cables are
plugged into device
AC Line Filters, Use of
Recommended UPS system at
installation site .
Charge battery, carry device while
on testing screen around care unit,
moving towards and away as noise
amplifies on screen to target EMI
offender – Note: Sources of EMI
may still emit interfering signals
when placed in stand by mode .
Recommendation is that identified
sources of EMI are either removed
from the testing environment,
testing environment is moved or
devices are fully powered down
and unplugged from AC power .
i . Top Offenders:
ii . Ultrasound Machines . See
“Recommended Separation
Distances” for RF interference
guidance on page 113
iii . Elevator or Lift Electric Motors
iV . XRay
V . Centrifuges
vi . Personal Electronics on
Patient’s Person
Note: If you are experiencing an issue with the device’s functionality, please capture export
logs (see page 88) and email logs to [email protected].