| MyoVista® Wavelet ECG (
wav
ECG™) 12-Lead Cardiac Testing Device
Appendix A: Technical Data
EMC Declaration Tables
The MyoVista Device is intended for use in the electromagnetic environment specified below .
Electromagnetic Emissions
Emission test
Compliance
Electromagnetic environment - guidance
RF emissions
CISPR 11
Group 1
The MyoVista Device uses RF energy only for its internal function .
Therefore, its RF emissions are very low and are not likely to cause
any interference in nearby electronic equipment .
RF emissions
CISPR 11
Class A
The MyoVista Device is suitable for use in commercial and hospital
quality establishments .
Harmonic emissions IEC
61000-3-2
Class A
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Complies
Electronic Immunity
Immunity test
IEC 60601
test level
Compliance level
Compliance level
Electrostatic
discharge (ESD)
IEC 61000-4-2
±8 kV contact
±15 kV air
±8 kV contact
±15 kV air
Floors should be wood, concrete or ceramic tile . If floors
are covered with synthetic material, the relative humidity
should be at least 30% .
Electrical fast
transient/burst
IEC 6100-4-4
±2 kV for power
supply lines
±2 kV for input/
output lines
±2 kV for power
supply lines
±2 kV for input/
output lines
Mains power quality should be typical of a commercial or
hospital environment .
Surge
IEC 61000-4-5
±1 kV differential
mode
±2 kV common
mode
±1 kV differential
mode
±2 kV common
mode
Mains power quality should be typical of a commercial or
hospital environment .
1Voltage
dips, short
interruptions and
voltage variations
on power supply
input lines
IEC 61000-4-11
<5% voltage for
0 .5 cycles; <5%
voltage for 1 cycle;
70% voltage for
25 cycles; <5%
voltage for 250
cycles
<5% voltage for
0 .5 cycles; <5%
voltage for 1 cycle;
70% voltage for
25 cycles; <5%
voltage for 250
cycles
Mains power quality should be typical of a commercial
or hospital environment . The MyoVista Device provides
an internal battery for uninterrupted use during power
mains interruptions .
Power frequency
(50/60Hz)
magnetic field
IEC 61000-4-8
3 A/m
3 A/m
Power frequency magnetic fields should be at levels
characteristic of a typical location in a typical commercial
or hospital environment .