USER MANUAL DO-LBL-06035(L)
| MyoVista® Wavelet ECG (
wav
ECG™) 12-Lead Cardiac Testing Device
Accuracy (Amplitude)
Meets or exceeds ANSI/AAMI/EN/IEC 60601-2-25:2011
Frequency range
0 .05 - 200 Hz (IEC/AHA)
Accuracy (Timing)
Meets or exceeds ANSI/AAMI/EN/IEC 60601-2-25:2011
Measurement range
Dynamic ± 10 mV, DC ± 300 mV
CMRR
>60 dB
Input impedance
≥2 .5MΩ (actually arrived ≥50MΩ)
Defibrillation protection
5,000 VDC
Defibrillation recovery time
<5 s
ECG
Durations and isoelectric waves
From Glasgow Interpretive Analysis documentation, “With respect to global QRS
duration, the Glasgow program measures QRS duration from the global QRS onset
to the global QRS termination . This means that an isoelectric segment within one
particular QRS complex by definition will lead to a shorter QRS duration for that lead
compared to the global QRS duration .”
ECG Distortion test filters:
No device filters used or required during testing .
Appendix A: Technical Data