![HeartSciences MyoVista wavECG User Manual Download Page 49](http://html.mh-extra.com/html/heartsciences/myovista-wavecg/myovista-wavecg_user-manual_2114879049.webp)
40
USER MANUAL DO-LBL-06035(L)
| MyoVista® Wavelet ECG (
wav
ECG™) 12-Lead Cardiac Testing Device
Patient Testing
Electrode Colors
Table 5-1 . AHA Electrode Identification
System
Electrode ID
Color Code
Limb Leads
RA - Right Arm
White
LA - Left Arm
Black
RL - Right Leg
Green
LL - Left Leg
Red
Precordial Leads
V1 = 4th intercostal space to the right of the sternum
Red
V2 = 4th intercostal space to the left of the sternum
Yellow
V3 = Directly between the leads V2 and V4
Green
V4 = 5th intercostal space to the left of the sternum
Blue
V5 = Level with V4 at left anterior axillary line
Orange
V6 = Level with V5 at left midaxillary line (directly under the midpoint of
the armpit)
Violet
Table 5-2 . IEC Electrode Identification
System
Electrode ID
Color Code
Limb Leads
RA - Right Arm
Red
LA - Left Arm
Yellow
RL - Right Leg
Black
LL - Left Leg
Green
Precordial Leads
C1 = 4th intercostal space to the right of the sternum
Red
C2 = 4th intercostal space to the left of the sternum
Yellow
C3 = Directly between the leads V2 and V4
Green
C4 = 5th intercostal space at the midclavicular line
Brown
C5 = Level with V4 at left anterior axillary line
Orange
C6 = Level with V5 at left midaxillary line (directly under the midpoint of
the armpit)
Violet