![HeartSciences MyoVista wavECG User Manual Download Page 118](http://html.mh-extra.com/html/heartsciences/myovista-wavecg/myovista-wavecg_user-manual_2114879118.webp)
109
USER MANUAL DO-LBL-06035(L)
| MyoVista® Wavelet ECG (
wav
ECG™) 12-Lead Cardiac Testing Device
Appendix A: Technical Data
Environmental Specifications - MyoVista Center Post Electrodes
Specification
Description
MyoVista
wavECG Device & MyoVista Cable 10 Leads AHA/IEC
Temperature
Operating
+5°C to +40°C (41°F to 104°F)
Storage
-20°C to +50°C (-4°F to 122°F)
Humidity (non-condensing)
Operating
≥80% RH non-condensing
Storage
≥80% RH non-condensing
Center Post Electrodes
Temperature
Operating
+4°C to +38°C (40°F to 100°F)
Storage
+4°C to +38°C (40°F to 100°F)
Humidity (non-condensing)
Operating
≥80% RH non-condensing
Storage
≥80% RH non-condensing
MyoVista
wavECG System (Boxed)
Temperature
Storage
+4°C to +38°C (40°F to 100°F)
Humidity (non-condensing)
Storage
≤80% RH non-condensing
ECG Specifications
Specification
Description
ECG
Leads
Standard 12 simultaneous leads
Touchscreen Display
Leads
12-channel display of the selected leads, selectable amplitude 5, 10, 20 mm/mV
Status
Filter status (on/off) / Insufficient electrode contact / Date-time-Patient demographics
Printout
Chart print-out speed
5/25/50/100 mm/s (manual print)
Sensitivities
5/10/20 mm/mV, either automatically adjusted or manually selected
Recording track
12-channel presentation, automatic baseline adjustment
Data Record
Fields
Patient data (name, age, sex, height, weight, BSA), user ID, and logo with optional
interpretation program ECG measurements results
Filter
Muscle noise filter
Myoelectric filter: 25 Hz, 35 Hz, 45 Hz, 75 Hz, and 100 Hz can be switched on/off .
ECG Amplifier
Simultaneous recording
8 channels
Sampling frequency
1000 Hz
Resolution
<5 µV/14 bit