![HeartSciences MyoVista wavECG User Manual Download Page 43](http://html.mh-extra.com/html/heartsciences/myovista-wavecg/myovista-wavecg_user-manual_2114879043.webp)
34
USER MANUAL DO-LBL-06035(L)
| MyoVista® Wavelet ECG (
wav
ECG™) 12-Lead Cardiac Testing Device
MyoVista Operations
Using the Built-in Battery
Due to storage and transportation time, the initial battery charge might not be sufficient to power
up the device or support a complete patient cycle . In order to provide sufficient charge to the device,
all new units should be charged for 8 hours upon receipt . When using the device for the first time,
operate the MyoVista Device on AC power only . Before running a test, always confirm that the battery
has sufficient power to complete the test .
Caution: Do not store new (unpackaged, unused) devices more than 6 months as the battery
may become inoperable if stored past 6 months without being charged.
Note: HeartSciences recommends keeping the device plugged into AC power whenever an AC
power outlet is readily available.
Monitoring the Battery
Monitor the battery gauge on the status bar of the device . If the battery begins to deplete at a faster rate
than usual, it may be time to replace the battery . Contact an authorized representative for service .
Maximizing the Battery Life
Use:
Use device on AC power unless it is being transported or used briefly in a location where an AC
power outlet is not readily available .
Not In Use:
While not in use, store the device plugged into an AC outlet, with the device turned ON
with the display active and the battery charge icon showing the device is charging . Alternatively, after
booting up the device to the Home Screen on AC power, leave it plugged into AC power and turn the
device OFF . This ensures the device is still receiving power to charge the battery from the AC charger
while in the OFF state .
If devices are stored long term, please schedule time to charge devices for at least 8 hours every 6
months .
HeartSciences recommends charging devices to 100% within 30 days of receiving a unit or prior to
long term storage .