Chapter 8 Equipment cleaning and maintenance
8.1 Repair and maintenance
If the equipment or accessories are damaged, or their original functions can’t be guaranteed,
discontinue use of the equipment.
If you need to have the equipment repaired, cleaned, preventively checked or maintained, please
contact your equipment provider or manufacturer (Micomme Medical)
8.2 Frequency
Based on standard hygiene considerations, the manufacturer recommends the following
maintenance schedule:
Clean the equipment, mask and hose before the initial use, then weekly;
Empty any remaining water in the water tank each morning and clean the tank
thoroughly at least once per week to prevent the growth of bacteria;
Check and clean the air filter once a month and replace it every six months;
Replace the face mask and hose every 6-12 months or as required.
8.3 Cleaning method
8.3.1 Cleaning of CPAP machine and hose
Disconnect the equipment from the power supply before cleaning it or carrying out regular
maintenance.
Clean the front panel and exterior enclosure with soft damp cloth, moistened with warm water or
mild disinfectant or CPAP cleaning wipes. Dry the equipment completely before inserting the
power connection.