Chapter 7 Alerts
7.1 Visual and Audible Alerts
If an unexpected fault event occurs with the Machine during use, an alert message in red text
will be displayed in the upper left corner of the screen, accompanied by an audible beeping
sound.
If power is disconnected from the Machine during use (Eg. A power cut), 2 beeps will sound,
followed by a 15 second pause. This will continue until power has been restored, or until the
internal device power has been exhausted.
7.2 Alert messages
Name
Alert description
Alert principle
Handling measures
Turbine
The turbine blower /
motor stops working
A sensor detects rotation
of the turbine rotor. If
rotation is not detected
within a certain period of
time, an alert will be given.
Immediately disconnect
from the power and contact
your service provider.
Heating
humidifier
The temperature in
heating humidifier
becomes too high
When the temperature of
heater element is over
90
℃
for more than 10
seconds, an alert will be
given.
Immediately disconnect
from the power and contact
your service provider.
Humidifier
fault
Humidifier is
disconnected
If the machine is to be
used without the
humidifier connected, it
Disconnect the power, wait
for 30 seconds. Before
trying again, turn the