User Manual
MK/YF38-07 – V2.01CS
© 2019 All Rights Reserved
Page 40 of 60
Disconnecti
on
The mask or tubing
hose is
disconnected from
the CPAP Machine
The pressure sensor
detects a problem for over
30 seconds, an alert will
be given.
Check if there is any
damage to the hose or face
mask; ensure the
connection between the air
outlet and hose is normal
and the face mask is
correctly worn; Mask leaks
should be under 60l/m
if the problem persists,
contact your service
provider.
Tubing block
The flow in the air
outlet tubing
becomes nearly 0
due to
obstruction/blockage
The flow sensor reads an
incorrect value for more
than 30 seconds, an alert
will be given.
Check if the hose or face
mask is blocked; ensure the
hose is not crushed or bent.
If problem persists, contact
your service provider.
7.3 Alert Suspension
If an alert is shown or heard, press the dial button during the alert to suspend it.
If the problem is not rectified within 30 seconds, the alert reminder will reoccur.
If you cannot resolve the alert, please contact your equipment provider for assistance.