User Manual
MK/YF38-07 – V2.01CS
© 2019 All Rights Reserved
Page 31 of 60
easier it will be to breathe out
against the airflow.
Ramp time
OFF or 5~45
minutes
Typically used in CPAP Mode,
the Ramp feature allows the
Machine to start at a lower
pressure and slowly ramp up to
the treatment pressure over the
Ramp time selected.
Patient Mode
Clinician Mode
Ramp start
4cmH2O ~ CPAP
Pressure
4cmH20 ~ min
CPAP pressure
When a Ramp time has been
selected, choose the starting
pressure. Over the Ramp time
selected, the Machine pressure
will slowly rise until;
a) The CPAP Treatment
Pressure has been reached
(CPAP Mode only)
or
b) The min CPAP pressure has
been reached (Auto CPAP Mode
only)
Stop and restart the Machine to
begin the Ramp time again.
Patient Mode
Clinician Mode
Humidifier
OFF. 1,2,3,4,5
Select the desired setting based
on
personal
comfort
and
environmental conditions.
The higher the number, the more
Patient Mode
Clinician Mode