User Manual
MK/YF38-07 – V2.01CS
© 2019 All Rights Reserved
Page 54 of 60
Table 2
Guidance and manufacturer’s declaration - electromagnetic Immunity
Immunity Test
IEC 60601-1-2
Test level
Compliance level
Electrostatic discharge
(ESD)
IEC 61000-4-2
±8 kV contact
±2 kV, ±4 kV, ±8 kV, ±15 kV air
±8 kV contact
±2 kV, ±4 kV, ±8 kV, ±15 kV air
Electrical fast
transient/burst
IEC 61000-4-4
±2 kV for power supply lines
±1 kV signal input/output
100 kHz repetition frequency
±2 kV for power supply lines
Not Applicable
100 kHz repetition frequency
Surge
IEC 61000-4-5
±0.5 kV, ±1 kV differential
mode
±0.5 kV, ±1 kV, ±2 kV common
mode
±0.5 kV, ±1 kV differential mode
Not Applicable
Voltage dips, short
interruptions and voltage
variations on power supply
input lines
IEC 61000-4-11
0 % UT; 0,5 cycle. At 0°, 45°,
90°, 135°, 180°, 225°, 270°
and 315°.
0 % UT; 1 cycle and 70 % UT;
25/30 cycles; Single phase: at
0°.
0 % UT; 250/300 cycle
0 % UT; 0,5 cycle. At 0°, 45°, 90°, 135°,
180°, 225°, 270° and 315°.
0 % UT; 1 cycle and 70 % UT; 25/30
cycles; Single phase: at 0°.
0 % UT; 250/300 cycle
Power frequency magnetic
field
IEC 61000-4-8
30 A/m
50Hz/60Hz
30 A/m
50Hz/60Hz