User Manual
MK/YF38-07 – V2.01CS
© 2019 All Rights Reserved
Page 39 of 60
must be switched off
electronically first,
otherwise an alert will be
given after 40 seconds.
humidifier off in the [Set] >
[Param] menu.
If the problem persists,
contact your service
provider.
Pressure
sensor
The pressure sensor
fails
If the pressure sensor
reads a negative value for
30 seconds or more, an
alert will be given.
Disconnect the power
supply, wait for 1 minute,
then re-apply the power and
run the Machine again. If
the problem persists,
immediately disconnect
from the power and contact
your service provider.
Flow sensor The flow sensor fails
The flow sensor reads
error data for over 30
seconds, an alert will be
given.
Disconnect the power
supply, wait for 1 minute,
then re-apply the power and
run the Machine again. If
the problem persists,
immediately disconnect
from the power and contact
your service provider.
High-
pressure
The air flow output
pressure exceeds
35cmH20
If the pressure sensor
detects the hose pressure
is over 35cmH20 for 6
seconds or more, an alert
will be given.
Immediately disconnect
from the power and contact
your service provider.