frequent Air Bubble alarms.
I)
In normal conditions, try to use AC or DC power source to extend battery service life. When use AC
power source, making sure it is well connected to ground and please use the power cord that is
standard configuration with the Infusion Pump. Just use battery when there is difficulty in ground
connection or without AC power (such as AC power failure or mobile infusion).
J)
Please do not use infusion pipe in the same pumping position for more than 6 hours (infusion pipe
replacement cycle for infusion at flow velocity below 1ml/h should be less than 24h). The infusion
pipe will deform and thus leading to flow error after long time use. It is suggested to move the infusion
pipe upward for 15cm every time after 6h (infusion pipe replacement cycle for infusion at flow velocity
below 1ml/h should be less than 24h) and then start infusion again, or replace with a new infusion
pipe.
K)
To prevent free flow after door open please make sure to close the flow clip of IV set before taking it
out of the Infusion Pump.
L)
Pay more attention to occlusion when infusion at low rate. The lower the rate, the more time needed
for detecting occlusion, thus there may be a long interval of infusion interruption.
M)
When using computer port, it may suffer interference from devices such as electric cautery apparatus,
mobile phone, wireless device or cardiac defibrillator etc. Please try to keep away from the
above-mentioned devices.
N)
If Infusion Pump falling off or suffering collision, stop using it immediately and contact the distributor or
the manufacturer. Even there is no damage on appearance or no malfunction alarm, the internal parts
may have damaged.
O)
The Infusion Pump must be operated by well-trained professionals such as doctor, nurse and medical
device expert.
P)
Do not disassemble or modify the Infusion Pump or use it for other purposes other than normal
infusion. Otherwise, the manufacturer takes no responsibility.
2. Introduction
2.1 Features
User friendly interface, easy operation
8 occlusion level adjustable