Drip rate alarm
1.Liquid Volume Empty alarm
Press SILENCE key to eliminate alarm
sound; detect if the infusion bottle is
empty and detect if the drop sensor is
correctly installed. Press Clear key and
then Start key to start infusion again.
Please ask the manufacturer to inspect
and repair if the alarm still cannot be
eliminated.
2. Abnormal drip rate
Press SILENCE key to eliminate alarm
sound and test if the infusion pipe is
blocked. Press Clear key and then Start
key to start infusion again. Please ask
the manufacturer to inspect and repair if
the alarm still cannot be eliminated.
Note:
The set parameters (including alarm setting) are stored in memory chips. The saved data will not be lost
after power failure and can be saved for more than 10 years. Alarm setting will be called out when the
machine is turned on again.
Method to verify the normal operation of alarm system:
Note: The verification below cannot be carried out when the infusion pipe is connected to patients;
otherwise, patients will be risky.
1). AC power-down alarm
Unplug AC power cord after turning the pump on, the device will display AC fails and give alarm
sound.
2). Operation forgotten alarm
The device will display “No Operation” and give alarm sound if no operation is made after 2min when
the infusion pump is in non-operation and non-alarm status.
3). Low battery alarm
The device will display low battery and give alarm sound when battery voltage is lower than 7.1V.
4). Battery exhausted alarm