8.1.10 Replacement of infusion set and liquid container
★
If you need to replace IV set, please follow steps below:
Close the flow clip of IV set. Open the pump door and take out the IV set.
As per instructions of 8.1.3, fill the new IV set with medical fluid and install it properly. Restart
infusion as required.
★
The IV set may be out of shape due to long-hours squeeze by the peristaltic system and which can
cause accuracy error. It is suggested that change the section of the infusion set that is against
peristaltic chips or replace a new infusion set after continuously working for 6 hours.
★
If need to replace infusion bottle, please follow steps below:
Close the flow clip of IV set. Open the pump door and take out the IV set.
Disconnect IV set from infusion bottle
Reconnect the IV set to a new infusion bottle.
Fill in and install the IV set as per instructions of 8.1.3.
Restart infusion as per infusion instructions of 8.1.
8.1.11 battery on indicator
1. When battery is supplying power, the screen will display “on battery ” and the battery indicator light
will be on.
2. When battery voltage is lower than 7.4V, the device will give an alarm and show low battery voltage.
3. In case of Li battery damaged or cannot be used, it should be replaced by professional personnel.
Wasted and old Li batteries cannot be disposed at will; instead, they should be sorted and then
disposed in order to prevent environmental from pollution.
8.2 Alarms and Solutions
The following alarms may occur in infusion preparation and infusion process. Please operate according
to the instructions.
(Table 3)
(See Table 1, 2 and 3 of Appendix 1 for alarm relevant parameters)
Name of alarms
Cause for alarms
Solutions