IV Set
BOON
Near Done
OFF
Drop/ml
1ml=20 drops
Night mode
OFF
Key lock
Never
Up pressure
Low
Calibration
Rate: 150ml/h
VTBI: 10ml
Dedicated IV set
OFF
Wireless
OFF
Drip sensor
OFF
Note: restore the factory settings, please calibrate the IV set again.
8.4.16.7 Drop/ml setting
After entering “D./ml” setting interface, press OK then the editable value is in white shade, press
or
to input requested drop/ml. it can be set between 0~60drop/ml.
When using external Drop Sensor, please follow the steps: Settings
Drop Rate settings
Drop
Sensor
ON. If the Drop Sensor is OFF, then means the Drop Sensor is disable. In Drip Mode,
though the unit is Drop/Minute, still run as flow rate, the formula as below:
Flow rate (ml/h)= Drip rate(D/min) * 60(min/h)* Drops(D/ml)
8.4.16.8 Air bubble level setting
Press OK to enter “Air Level” setting interface, set required air bubble size for detection(OFF, Lev1,
Lev2, Lev3), then press OK to save and exit.
Air bubble size for detection and value limit are shown as follows:
Air bubble size level
Value limit
Level 1
About 50ul
Level 2
About 100ul
Level 3
About 250ul
Level 4
About 500ul