of Air Bubble alarm, it is convenient for the user to clamp the flow clip and then squeeze the air
bubble back into the drip chamber.
H)
Fix the Infusion Pump well to infusion stand/bar and also ensure the stability of the stand/bar. Be
cautious when moving the stand/bar and the Infusion Pump to prevent the Infusion Pump falling off or
the stand collision with surrounding objects.
I)
The Infusion Pump cannot parallel use with gravity infusion device, as the machine can’t detect
downstream occlusion or empty of gravity infusion set.
J)
The Infusion Pump cannot use with possible large negative or positive pressure piping such as
extracorporeal circuit. As in such case, the Infusion Pump cannot ensure infusion accuracy and
correct alarm functions.
K)
The Infusion Pump cannot use for blood transfusion.
L)
Please install the IV set in correct direction (from left to right). If installing in a wrong direction, patient’s
blood may be sucked out.
M)
Do not use the Infusion Pump near inflammable liquid or gas.
N)
Never store or use this infusion pump in places with chemical and active gas (including gases for
disinfection) or with much humidity, since it may affect the internal parts of infusion pump and finally
lead to degradation or damage of their performance.
O)
This infusion pump cannot be directly powered by vehicle power supply.
P)
Medical staffs should have regular patrol inspection, instead of relying on alarm system only, so as to
prevent accidents.
Q)
Under single failure state, the maximum infusion quantity should be the remaining fluid in infusion
bottle/bag. Keep monitoring the operation state of system during infusion and check infusion pipe,
instead of relying on alarm function of this system.
R)
Operator should not start infusion until the infusion parameters of infusion pump are conforming to the
medical advices. Any nonconforming setting of parameters may lead to invalid operation.
S)
Make sure there is no water on the external wall of infusion pipe, especially make sure do not let water
flow to ultrasonic sensor; otherwise, it may fail to give alarm when there is bubble.
Cautions
Personal injury or property loss may occur if violating these precautions.
A)
Inspect the Infusion Pump before use, making sure it can work normally. If any malfunction is found,