28
(Fig. 14
TPN Mode
)
After entering parameter setting interface of
TPN mode
, press
to choose
parameter item to be edited, it will be marked with blue rectangle background. Press
OK key, the blue rectangle disappears, then the current editable position will be
marked white rectangle background, input the value according to (
Fig. 9
), VTBI, Up
time, Line time, Down time are adjustable, after confirming, press OK to save.
Rate will be automatically calculated by system according to VTBI, Up time, Line time
and Down time, and shows on the screen.
Return:
If want to reselect infusion mode, press
to return main menu (
Fig. 9
).
(
7
)
Drip Mode
(Fig. 15
Drip Mode
)
After entering the dripmode parameter setting interface, press
to select the
parameter item to be edited. The parameter item description with a blue rectangle
background frame is waiting for edited, press the OK key, the blue rectangle
background frame disappears, and the current editable bit number of this parameter
item appears white rectangle background frame. Set it according to (Figure 11:
Numerical Input Method). The drip rate and preset are adjustable. After confirmation,
press OK key to save.
Note: The drip mode must first according to the relevant statement of the infusion set
used, and the drip rate corresponding to the 1ml infusion volume should be entered.
(For example: the Boon infusion tube statement: 20 drops of distilled water is
equivalent to 1 ml
±
0.1 ml, then in the drop / ml parameter settings, enter: 20.)