42
will run at Bolus rate until finishing the VTBI, then run at original rate.
Press
to turn to next page.
8.4.6 Back light Setting
There are 10 levels of backlight. Select the progress bar per requirement, then press OK to save it. The
fewer bars, the darker backlight
8.4.7Key Sound Setting
After entering “Key sound” setting interface, select “ON” or OFF”, then press OK key to save and exit.
8.4.8 Forget Operate Alarm Setting
After entering “Forget Op” setting interface, select “ON” or “OFF”. When the setting is on: in ‘stop’ status,
the reminder alarm shall sound when no operation on keypad in 2 minutes.
8.4.9 Near Done Alarm Setting
After entering “Near Done” setting interface, select “ON” or “OFF”. If the setting is “ON” and with preset
VTBI, “Near Done” alarm should sound 3 minutes before VTBI is complete.
8.4.10 Infusion Log Setting
After entering “Infusion Log” setting interface, press OK to upload/view logs. Infusion pump could store
about 65000 pieces history records, including turn on/off, start, stop and alarms during operation. When
AC power failure or power off, the records could still be saved. When records are full, the new record
will overwrite the old one. The records won’t be lost even the device without power. The content of the
records cannot be modified.
Select “Upload log
‟
, all infusion records can be viewed on computer (only available when connect the
pump to computer by USB interface). Select “View log
‟
, the pump can directly display the latest
infusion/alarm information. Press RETURN key, the pump shall return to “parameter setting interface
‟
.
(1)
Upload log: upload infusion records to computer with USB cable. Please refer to steps as follows:
a. Open the super terminal (if there is no, please open the computer Main menu - Program - Annex -
Communication - Super terminal, establish the super terminal named 115200 on the desktop. Click
disconnect icon to disconnect terminal connection. In the file menu, select "properties", set the COM