9
Internal battery
Li_Polymer 7.4V 1900mAh;
Recharge time: 10h with power on, 3.5h with power off.
Running time: more than 10h at rate of 25ml/h, environment
temperature 25
℃
after being fully charged.
Charging and discharging cycle: more than 300 cycles
DC
DC12 ± 1.2V, 1.5A, CE certified
Power consumption
25VA
Fuse
Slow fuse Specification: 250V 2A
(Maximum fusing time is 10 seconds when current
is 5.5A)
Dimensions
212*138*80 (L*W*H )mm (Pole clamp excluded)
Weight
1.4Kg
5. Installation and Adjustment
5.1 Installation conditions and technical requirements
The Infusion Pump can be fixed to a vertical IV pole with diameter of 12-35mm, or on platform with
slope angle not exceeding 5°.
5.2 Installation method and cautions
Method 1: Put the infusion pump on stable platform
Method 2: Fix the infusion pump to IV Pole as per below steps:
(
1
)
Rotate the clamp knob of fixation screw out the rod, leave space for IV pole
(
2
)
Clamp to IV pole (the IV pole should meet the requirements of balance and mechanical strength),
screw down the knob to fix the position of infusion pump. Should hold the infusion pump during
the installation; only release the hand after screwed tightly to avoid falling.
Remark: Mechanical strength of IV pole: yield strength≥170MPa, tensile strength≥480MPa,
load-bearing requirement≥13kg , verticality of IV pole≤5°
。