21
(Fig. 5: Installation diagram of Hawkmed dedicated infusion set)
Attention: Please make sure there is no liquid on the external wall of infusion set, and no liquid flowing to
ultrasound sensor; otherwise, it may fail to give air bubble alarm when there is air in line.
Drop sensor connection:
1.
Install the drop sensor as per above figures (Note: The liquid level should be lower than the
red line label).
2.
Connect the drop sensor cable to DROP interface on back of machine.
8.1.4 Infusion parameters setting
Press OK key to select “InfuMode” in the main menu, the machine will display infusion modes: rate
mode, drip mode, time mode, weight mode, intermittent mode & link mode.