Frequent Air Bubble
alarm
The IV set is too soft or too thin
Replace IV set
Small air bubble in the IV set
Select a higher level air bubble
filter.
★
Note: After changed pressure
level, please observe patients’
condition continually, if feel
uncomfortable, shall stop
infusion and return to previous
pressure level immediately.
Accuracy discrepancy
The IV set is not calibrated
Calibrate the accuracy of IV set
The IV set currently used does not
match the default brand.
Select correct brand of IV set
Due to variation in weather and
temperature, the internal
parameters of the pump
incompatible with that of the IV set
actually used.
Re-calibrate the accuracy of IV
set.
Certain parts of the machine may be
defective.
Contact distributor or
manufacturer for repair
Beside the problems mentioned in 8.2, please contact the sales agent / manufacturer for repair.