35
3. The sensor of infusion pump
is faulted
Please contact the distributor or
manufacturer to repair.
AC powder-down
alarm
AC power cord falls off/is cut off
after turning the device on
Press SILENCE to eliminate alarm.
Insert the AC/AC power cord again.
Battery on alarm
1. AC/DC is not connected to
the infusion pump before turning
it on.
Press SILENCE key to eliminate alarm.
Check if the power line is plugged or
tightly plugged or not.
2. Power circuit of the infusion
pump is with problem.
Ask the manufacturer to inspect and
repair.
Low battery alarm
(battery can be
used only when
power failure or in
moving process)
1. At least 30min before the
infusion stops because of the
battery running out.
Press SILENCE key to eliminate alarm
sound. If the alarm is given after 2min
without AC connected, it will alarm
again. Stop using it and charge the
battery fully.
2. The battery is aging or the
charging circuit of infusion pump
is faulted.
Please contact the distributor or
manufacturer to repair.
Battery runningout
alarm (In cases
when only battery
can be used at
power failure or in
moving process)
1. At least 3min before the pump
turns itself off automatically for
the battery running out.
Stop using and charge the battery fully.
2. The battery is aging or the
charging circuit of infusion pump
is faulted.
Please contact the distributor or
manufacturer to repair.
AC fail alarm
when AC fail and battery exhaus
ted during infusion, the pump wil
l alarm in buzzing and flash red l
ight for 3 mins
Please charge it.